AIMTEC |
Click on Series name for product info on aimtec.com
Series AME40-MAZ
40 Watt | AC-DC / DC-DC Converter
Models
Single output
Model
AME40-3.3SMAZ
AME40-5SMAZ
AME40-12SMAZ
AME40-15SMAZ
AME40-24SMAZ
FEATURES:
4000VAC I/O isolation
Medical Switching power modules for PCB mounting
Operating temperature: -40 to +85°C
Universal Input: 90-264VAC, 47-440Hz, or 130-370VDC
Energy star compliant
RoHS compliant
Low Ripple and Noise
CE, cULus approvals
CB
Input Voltage
(VAC/Hz)
90-264/47-440
90-264/47-440
90-264/47-440
90-264/47-440
90-264/47-440
Input Voltage
(VDC)
130-370
130-370
130-370
130-370
130-370
Output
Voltage
(V)
3.3
5
12
15
24
Maximum
Capacitive
load (µF)
8200
8200
2200
1500
560
Output Current
max
(A)
8
8
3.33
2.66
1.66
Efficiency
(%)
115 230
VAC
VAC
77 75
80 80
84 85
84 84
82 82
Models
Dual output
Model
Input Voltage
(VAC/Hz)
Input Voltage
(VDC)
Output
Voltage
(V)
Maximum
Capacitive
load (µF)
Output Current
max
(A)
Efficiency
(%)
115 230
VAC
VAC
AME40-5DMAZ
90-264/47-440
130-370
±5 3300
±4 80 80
AME40-12DMAZ
90-264/47-440
130-370
±12
2200
±1.66
82 80
AME40-15DMAZ
90-264/47-440
130-370
±15
1000
±1.33
82 81
AME40-24DMAZ
90-260/47-440
130-370
±24
470 ±0.835
79 77
NOTE: All specifications in this datasheet are measured at an ambient temperature of 25°C, humidity<75%, nominal input voltage and at
rated output load unless otherwise specified.
Input Specifications
Parameters
Current
Inrush current <2ms
Leakage current
External fuse
Input dissipation
Start-up time
Conditions
115 VAC
230 VAC
115 VAC
230 VAC
Recommended slow blow type
115 / 230 VAC
Typical
3.15
0.5
150
Maximum
1000
530
30
50
0.15
Units
mA
mA
A
A
mA
A
W
ms
Output Specifications
Parameters
Voltage accuracy
Line regulation
Load regulation
Cross regulation
Transient response deviation
Ripple & Noise*
Hold-up time (min)
Minimum Load Current
Conditions
(LL-HL)
0-100% load single
0-100% load dual
25% load - 1st out, 100% load – 2nd out
25% load Step
3.3 / 5V models
12 / 15V models
24V models
115VAC
*Ripple & Noise measured at 20MHz bandwidth with 0.1μF M/C and 47μF E/C
Typical
±2
±2
±3
±3
±5
±2
75
100
150
15
0
www.aimtec.com
F 052e R8.K
Tel: +1 514 620 2722
1 of 4
Maximum
Units
%
%
%
%
% of Vout
mV p-p
ms
% of Max
Toll free: + 1 888 9 AIMTEC
(924 6832)
North America only
Series AME40-MAZ
40 Watt | AC-DC / DC-DC Converter
Isolation Specifications
Parameters
Tested I/O voltage
Isolation Resistance
Conditions
60 sec
Typical
>1000
Rated
4000
Units
VAC
MΩ
General Specifications
Parameters
Switching frequency
Protection class
Over current protection
Over voltage protection
Short circuit protection
Operating temperature
Maximum case temperature
Storage temperature
Temperature coefficient
Cooling
Humidity
Case material
Weight
Dimensions (L X H X W)
MTBF
Conditions
Typical
Maximum
47
Class II
Foldback
110 140
Zener diode clamp
110
Auto recovery
With derating above 50ºC
-40 - +85
100
-40 to +95
0.02
Free air convection
Non condensing
95
Plastic (flammability to UL 94V-0)
300
3.56 x 2.48 x 1.18 inches 90.5 x 63 x 30 mm
> 800,000 hrs (MIL-HDBK -217F, t=+25 oC)/Full Load
> 200,000 hrs (MIL-HDBK -217F, t=at highest operating temperature)/Full Load
Units
KHz
%
%
oC
oC
oC
% / oC
% RH
g
Environment Approval
Test
Shock
Vibration
Parameters
Wave form
Acceleration amplitude
Bump duration
Converter operation
Number of bumps
Test mode
Displacement
Acceleration
Converter operation
Conditions
Half sine wave
5gn
30 ms
Before and after test, body mounted (on chassis)
18 (3 in each direction for every axis)
Sweep sine, 10-100Hz, speed 0.05Hz/s
1 mm
3g, 3 loops 30min one cycle, 3h total, every axis tested
Before and after test, body mounted (on chassis)
Safety Specifications
Parameters
Agency approvals
cULus, CE, CB
Medical Electrical Equipment
Information technology Equipment
Harmonic Current Emissions
EMI - Conducted and radiated emission
Voltage fluctuations and flicker
Standards
Electrostatic Discharge Immunity
RF, Electromagnetic Field Immunity
Electrical Fast Transient/Burst Immunity
Surge Immunity
RF, Conducted Disturbance Immunity
Power frequency Magnetic Field Immunity
Voltage dips, Short Interruptions Immunity
IEC\EN\UL 60601-1, 2 x MOOP, CSA-C22.2 No. 601.1-M90
EN 60950-1:2006+A11:2009
IEC/EN 61000-3-2, Class A
EN55011, class B
IEC/EN 61000-3-3, (EN60555-3)
IEC 61000-4-2 Level 3
IEC 61000-4-3 Level 2
IEC 61000-4-4 Level 3
IEC 61000-4-5 Level 2
IEC 61000-4-6 Level 2
IEC 61000-4-8 Level 2
IEC 61000-4-11
www.aimtec.com
F 052e R8.K
Tel: +1 514 620 2722
2 of 4
Toll free: + 1 888 9 AIMTEC
(924 6832)
North America only
|